TLV-S Series

Vertical laboratory autoclaves for the sterilization of solid and liquid loads
Built-in video tutorials
7-inch capacitive touchscreen display
Push-button operated mechanically assisted door

Request a quote

Request a quote

Smart lab solutions

At LabSPX, we are committed to bringing innovative and efficient solutions to laboratories. Our Smart Lab Solutions are designed to streamline processes, enhance accuracy, and boost productivity.

Functionalities and technical specifications

Specification More info
Liquids and culture mediaSterilize your liquid loads professionally using Fâ‚€ and a flexible probe.
GlasswareEnsure the purity of your glass instruments with a steam generator that produces high-quality steam, minimizing salt transfer during processing.
Laboratory waste bagsWith the initial prevacuum pulse and steam generator, you can effectively sterilize waste bags without issues.
Biohazardous wasteThe bacteriological filter at the air inlet and outlet of the chamber enables the safe processing of biohazardous loads.
Plastic and metal objectsSterilize plastics and metal objects efficiently with faster cycles, thanks to the steam generator.

Multipurpose vertical laboratory autoclaves

The TLV-S Series Top Line vertical laboratory autoclaves are suitable for sterilizing both solid and liquid loads in pharmaceutical, biotechnology, chemical, and research laboratories. These fully automatic autoclaves combine modern engineering with advanced digital functionality to support demanding laboratory sterilization workflows.

All models feature a modern design with a 7-inch capacitive touchscreen and cloud-based connectivity. They are developed to support laboratories operating under FDA, GMP, and GLP environments through advanced electronic records and data management capabilities.

What sterilization technology does the TLV-Series use?

The TLV-S Series incorporates a range of technologies that simplify operation while ensuring reliable sterilization performance.

Key features include:

  • Push-button-controlled, mechanically assisted doors

  • Integrated steam generator for instant steam production

  • Air inlet and outlet fitted with bacteriological filters

  • Built-in vacuum pump for initial prevacuum

  • Fâ‚€-controlled sterilization using integrated flexible probes

  • Automatic water filling from the network

  • Water-cooled direct discharge system

Modern design

A 7-inch capacitive touchscreen provides access to all operating parameters and simplifies day-to-day use.

The interface includes:

  • Real-time program monitoring

  • User administration controls

  • Access to protocol history and video tutorials

  • Built-in ticket system for maintenance requests

  • Heating jacket for final drying and water-cooling coil for fast cooling

  • Integrated air compressor and vacuum pump

  • Medical-grade casters for mobility

Cloud-based connectivity and advanced technical support

The TLV-S Series offers cloud-based connectivity for equipment management and technical assistance. Users can request technical support directly from the controller, share the screen through TeamViewer, and authorize remote equipment diagnostics.

The cloud platform also enables centralized management of multiple autoclaves, tailored reporting, multi-device notifications, and automatic traceability for audits. All data is protected with SSL encryption. The management software can also be integrated into a private server environment using Docker, Active Directory, or LIMS.

Applications of the TLV-S Series

The TLV-S Series is suitable for sterilizing a variety of laboratory materials, including:

  • Culture media and liquids
  • Plastics
  • Metal objects
  • Glassware
  • Laboratory waste bags
  • Biohazardous waste

Sterilization process

The sterilization cycle begins with a prevacuum phase, during which a single vacuum pulse removes the chamber air. The system then injects saturated steam until it reaches the required temperature.

Once the target temperature has been achieved, it is maintained for the predefined sterilization time. After the sterilization phase is complete, the cycle finishes with a natural cooling stage.

Available models:

TLV-50 TLV-75 TLV-110 TLV-150
Total/usable volume of the chamber (L) 58/56 83/81 124/118 169/155
Usable dimensions of the chamber (Ø x H mm) 400 x 450 400 x 650 500 x 600 500 x 850
External dimensions (L x D x H mm) 610 x 870 x 1060 610 x 870 x 1110 710 x 980 x 1160 710 x 980 x 1310
Loading height (mm) 815 865 915 1065
Net weight (Kg) 131 139 195 210
Power (W) 3600 3600 9000 9000
Voltage* (V) 230 230 400 400
Frequency (Hz) 50/60 50/60 50/60 50/60

 

What is the difference between a vertical and a horizontal autoclave?

Read everything you need to know about autoclaves on this page. 

Brochures

Download your brochure here!

Brochure TLV-S Series
A multipurpose vertical laboratory autoclave is a top‑loading sterilization system designed to process both solid and liquid laboratory loads. It is commonly used in pharmaceutical, biotechnology, and research environments where reliable, high‑capacity steam sterilization is required.
The TLV‑S Series vertical laboratory autoclaves are used for sterilizing solid and liquid loads in pharmaceutical, biotech, and research laboratories. They are designed to support routine and critical sterilization processes while ensuring full compliance with regulated laboratory standards.
F₀‑controlled sterilization is a method that calculates the lethality of the sterilization process based on time and temperature. In the TLV‑S Series, flexible probes measure product temperature directly, allowing precise sterilization control for liquid loads and improved process validation.
Yes, the TLV‑S Series is designed to comply with FDA, GMP, and GLP requirements by integrating secure electronic records, digital data control, and traceability features suitable for regulated laboratory environments.
A vertical autoclave is loaded from the top and is often preferred for liquid loads and space‑efficient installations, while a horizontal autoclave is front‑loading and typically used for higher throughput and larger batch processing.
A vertical laboratory autoclave sterilizes materials using saturated steam under controlled temperature and pressure conditions. The process typically includes a prevacuum phase, steam injection, a sterilization holding phase, and a cooling phase to complete the cycle safely.

The prevacuum phase removes air from the sterilization chamber before steam is introduced. Removing air improves steam penetration and helps achieve more uniform sterilization throughout the load.

Related instruments

Explore similar solutions and complementary products.

Vertical autoclaves without drying
AES Series
Vertical laboratory autoclaves without drying to sterilize laboratory materials
Cost-effective
Both solid and liquid laboratory loads
Plug-and-play installation
ahs n raypa 1
AHS-N Series
Front-loading benchtop autoclaves without drying for a wide range of solid and liquid loads
Cost-effective and high-performance
Suitable for laboratories with limited space
Plug-and-play installation
Vertical laboratory autoclaves with prevacuums and drying
AE-B Series
Vertical laboratory autoclaves with prevacuums and drying
Multiple types of sterilization
Entirely automated with a 5'' touchscreen
Solids loads dry at end of cycle