Calibration and quality control

Table of Contents

Calibration and quality control procedures ensure that a laboratory balance continues to provide reliable and traceable mass measurements over time. This cluster covers what calibration means, how it differs from adjustment, the tools used to perform it, and how balance performance is verified between full calibrations.

What is calibration?

Calibration is the process of comparing a balance’s readings against certified reference weights of known mass to determine whether the instrument measures accurately across its range. Calibration itself does not change or correct the balance. It only documents the deviation between the displayed reading against the certified value of the reference weight.

What is the difference between calibration and adjustment?

Calibration measures and records how far a balance’s readings deviate from true values, while adjustment is the corrective step that resets the balance’s internal parameters so its readings match the reference weights used. In practice, a calibration check is often followed by an adjustment if the balance falls outside its acceptable deviation, after which a second calibration check confirms the adjustment was successful.

  • Calibration: identifies deviation, does not alter the balance
  • Adjustment: corrects the balance’s internal parameters to bring readings back within acceptable limits
  • Both steps are typically performed together during routine balance maintenance

How does balance calibration work?

Balance calibration involves placing one or more certified reference weights on the pan and comparing the displayed reading to the weight’s certified mass. Depending on the procedure used, calibration may include tests at multiple points across the weighing range to evaluate linearity. The deviation at each point is recorded, and if it exceeds the balance’s tolerance, an adjustment is performed using either the same reference weights or the balance’s internal calibration mechanism.

What is the difference between internal and external calibration?

Internal calibration: uses a built-in motorized weight inside the balance that can be engaged automatically or on demand, allowing the balance to perform calibration automatically without manually placing external weights.

External calibration: requires an operator to manually place certified external weights on the pan in a defined sequence, offering direct traceability to the specific reference weights used.

Internal calibration is faster and reduces handling error, while external calibration allows verification against weights with documented calibration certificates from a specific source.

Calibration weights

Calibration weights are precision-manufactured reference masses used to calibrate and adjust laboratory balances. They are classified according to defined tolerance classes, such as E1, E2, F1, F2, and M1, with E1 representing the highest accuracy class and the smallest permissible tolerance. The appropriate weight class depends on the balance’s readability, required measurement uncertainty, and application requirements. Using calibration weights with insufficient accuracy can introduce additional uncertainty into the calibration process.

  • Available in individual weights or weight sets covering a range of values
  • Typically constructed from stainless steel to resist corrosion and mass drift
  • Require careful handling with forceps or gloves to prevent contamination that could affect their certified value

Balance verification

Balance verification is a routine check, typically performed daily or before critical weighing’s, that confirms that the balance remains within predefined performance limits. It is a faster, more frequent check intended to catch drift or malfunction between scheduled calibrations, rather than a replacement for full calibration and adjustment.

How often should a balance be calibrated?

Calibration frequency depends on how often the balance is used, the required measurement accuracy, the stability of the weighing environment, and the requirements of the applicable quality system. Many laboratories perform calibration at intervals ranging from several months to one year, combined with regular verification checks using reference weights between calibrations. Balances used for critical measurements, high-throughput weighing, or highly sensitive applications such as trace analysis may require more frequent calibration or verification due to increased sensitivity to environmental changes. A balance should also be calibrated or verified after relocation, servicing, or significant environmental changes, as these events can affect measurement performance.

Read more:

Explore our balances here