How to clean an autoclave?

Share

Regular cleaning and maintenance are essential to keep an autoclave operating safely and reliably. Over time, residues from sterilization cycles, including mineral deposits, biological material, and limescale, can accumulate inside the chamber and components, reducing sterilization effectiveness and posing safety risks to operators.

This guide covers which products to use, how to clean each part of the autoclave correctly, and how often each task should be performed.

Why is cleaning an autoclave important? 

The performance and safety of an autoclave depend directly on regular cleaning.

Without proper maintenance:

Autoclaves operate under high temperature and pressure, and residue buildup directly affects their performance. Deposits inside the sterilization chamber can interfere with steam distribution, reducing the reliability of sterilization results. Limescale on the gasket or lid can compromise the pressure seal, while blockages in the drainage filter can affect cycle completion and water management.

Beyond performance, neglected maintenance can shorten equipment lifespan and, in some cases, create safety risks for operators. Consistent cleaning is therefore not just a quality measure, it is a safety requirement.

Always consult the manufacturer’s manual

Before using any cleaning product or procedure, consult the user and maintenance manuals for your specific autoclave model. These documents specify approved cleaning products, routine maintenance procedures, and any technical requirements particular to that unit.

Following manufacturer guidelines helps preserve the warranty, ensures compliance with industry standards, and reduces the risk of accidental damage.

What products can you use to clean an autoclave?

Using the correct cleaning materials is essential to prevent damage and maintain performance.

Recommended cleaning products include:

  • Soft cloths or cotton towels
  • Distilled water to prevent mineral buildup
  • Neutral detergent or a non-abrasive cleaning product (as recommended by the manufacturer)
  • Protective gloves

Avoid:

  • Abrasive materials
  • Corrosive substances

What does autoclave cleaning involve?

Routine cleaning involves a series of simple manual steps that must be carried out carefully to prevent damage to the equipment. It’s important to observe general precautions such as:

  • Avoid using direct water jets, whether high-pressure or spray
  • Prevent the buildup of limescale or other residues in the sterilization chamber, lid, and gasket

Recommended cleaning procedures:

Step 1: clean the gasket, lid interior, and external surface

  • Use a clean cloth dampened with water to wipe the gasket and lid interior
  • Clean external surfaces with a soft cloth and, if needed, a small amount of neutral detergent
  • Dry all surfaces thoroughly before using the autoclave
  • External surfaces can occasionally be disinfected with denatured alcohol

Step 2: clean the sterilization chamber and accessories

  • Use a cloth dampened with water and, if necessary, a small amount of neutral detergent
  • Clean the chamber, heating element cover, baskets, and internal surfaces
  • Rinse thoroughly with distilled water to remove residue

Important:

  • Do not use sharp or pointed tools to remove limescale
  • If buildup is significant, check the quality of the water used

Step 3: clean the water reservoirs

For autoclaves with clean and wastewater tanks:

  • Empty and clean the tanks regularly

Important safety note:

  • Never empty tanks while the autoclave is under pressure
  • Hot water or steam can cause burns

Step 4: cleaning the drainage filter

Residues can accumulate in the drainage filter and cause blockages.

  • Remove the filter following the manufacturer’s instructions
  • Clean it under running water

If needed:

  • Use a pointed tool for stubborn debris
  • Use hot water if the filter is clogged with agar

If the filter remains blocked after cleaning, it should be replaced with an original spare part.

How to clean an autoclave chamber?

The sterilization chamber requires more careful cleaning than external surfaces, since residue or scale buildup directly affects steam distribution and cycle performance.

  • Wipe the chamber, heating element cover, and internal surfaces with a cloth dampened in water and a small amount of neutral detergent.
  • Remove baskets, trays, and racks before cleaning, and clean each separately.
  • For limescale deposits, use a descaling solution suited to autoclave chambers, such as a diluted citric acid solution, applied per the manufacturer’s instructions, never a chlorine-based or abrasive cleaner, which can pit or corrode stainless steel.
  • Never use steel wool, wire brushes, or sharp tools on the chamber walls, these scratch the surface and create sites for future corrosion.
  • Rinse all surfaces thoroughly with distilled water to remove detergent or descaling residue before the next cycle.
  • Dry the chamber and accessories completely, residual moisture promotes mineral deposits and spotting.
  • Inspect chamber walls for discoloration, pitting, or corrosion while cleaning, early signs are easier to address than advanced damage.
  • If limescale recurs frequently, check the hardness of the water supply, hard water is the most common cause of chamber scaling.

How often should an autoclave be cleaned?

The cleaning and maintenance schedule for an autoclave should be tailored to its workload and the types of sterilization cycles performed. 

Example cleaning schedule (always defer to the manufacturer’s instructions): 

Frequency Recommended maintenance
Daily Clean the gasket, wipe the interior of the lid, and clean external surfaces
Weekly Clean the sterilization chamber and accessories, and replace the water in the tanks if required
Monthly Clean the drainage filter and check the safety valve for correct operation
Periodically Descale the chamber, typically every one to three months depending on water hardness, and inspect door seals and gaskets for wear
As required Replace worn or damaged gaskets, decontaminate immediately after a spill or contamination event, and recalibrate temperature and pressure sensors

The exact cleaning frequency depends on the autoclave model, workload, water quality, and materials processed. Always follow the manufacturer’s maintenance schedule where it differs from these general recommendations.

Maintenance tips after cleaning an autoclave

Beyond routine cleaning, a few additional practices help maintain long-term performance. A consistent inspection checklist covers the following areas:

  • Gasket condition: inspect for wear, deformation, or brittleness, a damaged gasket compromises the pressure seal and should be replaced promptly.
  • Chamber condition: check for scale buildup, discoloration, or corrosion on interior walls, which can affect heat transfer and cycle efficiency.
  • Drainage filter: confirm the filter is free of debris and correctly seated, a blocked filter slows drainage and extends cycle times.
  • Water quality: use distilled or demineralized water where specified, hard or mineral-rich water accelerates scale formation and gasket wear.
  • Water reservoirs: check tank levels and replace water on schedule, stagnant or contaminated water affects steam quality.
  • Door and lid sealing: verify the door closes and locks correctly, a poor seal indicates gasket wear or misalignment and risks cycle failure.
  • Temperature and pressure readings: compare displayed values against expected cycle parameters, deviations can signal sensor drift or a developing fault.
  • Sensors: calibrate temperature and pressure sensors periodically to keep cycle parameters accurate and within specification.
  • Consumable and wear components: replace gaskets, filters, and seals with original spare parts to maintain compatibility and performance.
  • Maintenance records: log every cleaning, inspection, and part replacement, a documented history supports troubleshooting and demonstrates consistent upkeep.